@misc{10498/30106, year = {2007}, url = {http://hdl.handle.net/10498/30106}, abstract = {Micro-pattern gas detectors (MPGD) could become suitable devices to carry out time-resolved X-ray diffraction experiments in the sub-millisecond time scale at synchrotron radiation facilities. Nevertheless, these devices are not free of problems. Among other problems are the build-up of ions in the detector components and the susceptibility of the materials in the structure to dielectric breakdown. Here, we present a new MPGD design we have called micro reading mesh chamber (MRMC). Its layout is based on a resistive anode, support pillars and a mesh formed by two planes of pick-up strips. In order to prevent charging up, a common problem with GEMs and MPGD in general, a minimum amount of dielectric material has been used in the design. The aim has been to avoid dielectric near the drift paths of the ions. To prevent dielectric breakdown, a resistive anode with guard rings at the base of the pillars has been designed. Finally, to improve the cathode signals quality, we use a mesh to read the avalanche. In this manner the size of the induced charge is larger than in more conventional eometries.}, publisher = {ELSEVIER}, keywords = {X-ray detector; Gaseous detector; Position-sensitive detector; Sub-millisecond; MPGD; Synchrotron radiation}, keywords = {X-ray detector}, keywords = {Gaseous detector}, keywords = {Position-sensitive detector}, keywords = {Sub-millisecond}, keywords = {MPGD}, keywords = {Synchrotron radiation}, title = {New micro-pattern gas detector based on a 2D position readout mesh}, doi = {10.1016/j.nima.2006.10.395}, author = {Fernández Banqué, F. and Ramos Lerate, Inmaculada Concepción and Martínez, J.C. and Beltrán, D. and Bordas, J. and Key, M.}, }