INTERNATIONAL JOURNAL OF HYPERTHERMIA 2021, VOL. 38, NO. 1, 846–861Abstract
Purpose: Bearing partially or fully metallic passive implants represents an exclusion criterion for
patients undergoing a magnetic hyperthermia procedure, but there are no specific studies backing
this restrictive decision. This work assesses how the secondary magnetic field generated at the surface
of two common types of prostheses affects the safety and efficiency of magnetic hyperthermia treatments
of localized tumors. The paper also proposes the combination of a multi-criteria decision analysis
and a graphical representation of calculated data as an initial screening during the preclinical risk
assessment for each patient.
Materials and methods: Heating of a hip joint and a dental implant during the treatment of
prostate, colorectal and head and neck tumors have been assessed considering different external
field conditions and exposure times. The Maxwell equations including the secondary field
produced by metallic prostheses have been solved numerically in a discretized computable
human model. The heat exchange problem has been solved through a modified version of the
Pennes’ bioheat equation assuming a temperature dependency of blood perfusion and metabolic
heat, i.e. thermorregulation. The degree of risk has been assessed using a risk index with
parameters coming from custom graphs plotting the specific absorption rate (SAR) vs temperature
increase, and coefficients derived from a multi-criteria decision analysis performed following
the MACBETH approach.
Results: The comparison of two common biomaterials for passive implants - Ti6Al4V and CoCrMo
- shows that both specific absorption rate (SAR) and local temperature increase are found to be
higher for the hip prosthesis made by Ti6Al4V despite its lower electrical and thermal conductivity.
By tracking the time evolution of temperature upon field application, it has been established
that there is a 30 s delay between the time point for which the thermal equilibrium is reached at
prostheses and tissues. Likewise, damage may appear in those tissues adjacent to the prostheses
at initial stages of treatment, since recommended thermal thresholds are soon surpassed for
higher field intensities. However, it has also been found that under some operational conditions
the typical safety rule of staying below or attain a maximum temperature increase or SAR value
is met.
Conclusion: The current exclusion criterion for implant-bearing patients in magnetic hyperthermia
should be revised, since it may be too restrictive for a range of the typical field conditions used.
Systematic in silico treatment planning using the proposed methodology after a well-focused