Show simple item record

dc.contributor.authorRincón Casado, Alejandro 
dc.contributor.authorLarrodé Díaz, Mauricio
dc.contributor.authorFernández Zacarías, Francisco 
dc.contributor.authorHernández Molina, Ricardo 
dc.contributor.otherIngeniería Mecánica y Diseño Industriales_ES
dc.contributor.otherMáquinas y Motores Térmicoses_ES
dc.date.accessioned2021-11-17T11:37:22Z
dc.date.available2021-11-17T11:37:22Z
dc.date.issued2021-09
dc.identifier.issn1996-1073
dc.identifier.urihttp://hdl.handle.net/10498/25786
dc.description.abstractThis work describes the design, construction and testing of a thermo-electric conditioning device installed in a neonatal incubator with the aim of improving the precision in the regulation of the interior air temperature, reducing noise and interior vibration, and improving the life of the neonate. A simplified one-dimensional thermal model has been developed, made up of resistances and thermal capacities that simulate the thermal behaviour of all the elements of the system from end to end. All the equations of the model are obtained in a nodal way, allowing the mathematical relationship between the input and output to be known. This model makes it possible to improve temperature control, avoiding the deviations that occur in the traditional model controlled by sensors at both ends. The computational model allows to predict the variation of temperatures in transient and permanent regime. This model allows the design and sizing of the thermoelectric system for different outdoor environmental conditions and the selection of the number of Peltier modules needed to satisfy the heating demand of other incubators with different geometry and capacity. The results of the computational model show good agreement with the experimental tests, despite being a simplified 1D nodal model. The results obtained show a coefficient of operation (COP) of 1.38, achieving higher performance than the current traditional electrical resistance system (COP = 1). In addition, a CFD study has been carried out to check the air patterns, to see the temperature uniformity and to estimate the number of air changes per hour (HVAC) inside the incubator.es_ES
dc.description.sponsorshipThis work was funded by University of Cadiz.es_ES
dc.formatapplication/pdfes_ES
dc.language.isoenges_ES
dc.publisherMDPIes_ES
dc.rightsAtribución 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/*
dc.sourceEnergies 2021, 14(17), 5278es_ES
dc.subjectthermal comfortes_ES
dc.subjectneonatal incubatorses_ES
dc.subjectPeltier effectes_ES
dc.subjectdynamic simulationes_ES
dc.subjectexperimental testinges_ES
dc.subjectthermoelectric systemes_ES
dc.titleExperimental and Computational Model for a Neonatal Incubator with Thermoelectric Conditioning Systemes_ES
dc.typejournal articlees_ES
dc.rights.accessRightsopen accesses_ES
dc.identifier.doi10.3390/en14175278


Files in this item

This item appears in the following Collection(s)

Show simple item record

Atribución 4.0 Internacional
This work is under a Creative Commons License Atribución 4.0 Internacional