Show simple item record

dc.contributor.authorSierra Padilla, Alfonso 
dc.contributor.authorLópez-Iglesias, David
dc.contributor.authorCalatayud‑Macías, Paloma
dc.contributor.authorGarcía Guzmán, Juan José 
dc.contributor.authorPalacios Santander, José María 
dc.contributor.authorCubillana Aguilera, Laura 
dc.contributor.otherQuímica Analíticaes_ES
dc.date.accessioned2023-05-16T07:24:25Z
dc.date.available2023-05-16T07:24:25Z
dc.date.issued2023-04-04
dc.identifier.issn1436-5073
dc.identifier.urihttp://hdl.handle.net/10498/28549
dc.description.abstractA new electrochemical sensor device has been developed through the modification of a polyaniline-silicon oxide network with carbon black (CB). Enhanced electrical conductivity and antifouling properties have been achieved due to the integra- tion of this cheap nanomaterial into the bulk of the sensor. The structure of the developed material was characterized using Fourier transform infrared spectroscopy, energy-dispersive X-ray spectroscopy, and scanning electron microscopy tech- niques. Cyclic voltammetry was used to characterize electrochemically the Sonogel-Carbon/Carbon Black-PANI (SNG-C/ CB-PANI) sensor device. In addition, differential pulse voltammetry was employed to evaluate the analytical response of the sensor towards sundry chlorophenols, common environmental hazards in aqueous ecosystems. The modified sensor material showed excellent antifouling properties, which led to a better electroanalytical performance than the one displayed with the bare sensor. Notably, a sensitivity of 5.48 × 103 μA mM−1 cm−2 and a limit of detection of 0.83 μM were obtained in the determination of 4-chloro-3-methylphenol (PCMC) at a working potential of 0.78 V (vs. 3 M Ag/AgCl/KCl), along with proficient values of reproducibility and repeatability (relative standard deviation < 3%). Finally, the analysis of PCMC was carried out in multiple validated water samples using the synthesized SNG-C/CB-PANI sensor device, obtaining excellent results of recovery values (97–104%). The synergetic effect of polyaniline and carbon black leads to novel antifouling and electrocatalytic effects that improve the applicability of this sensor in sample analysis versus complex conventional devices.es_ES
dc.formatapplication/pdfes_ES
dc.language.isoenges_ES
dc.publisherSpringeres_ES
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.sourceMicrochimica Acta, Vol. 190, Núm. 5, 2023es_ES
dc.subjectConducting polymerses_ES
dc.subjectNanomaterialses_ES
dc.subjectSonogel-Carbon electrochemical sensorses_ES
dc.subjectDifferential pulse voltammetryes_ES
dc.subjectChlorophenolses_ES
dc.subjectAntifouling propertieses_ES
dc.subjectHigh-energy ultrasoundes_ES
dc.titleIncorporation of carbon black into a sonogel matrix: improving antifouling properties of a conducting polymer ceramic nanocompositees_ES
dc.typejournal articlees_ES
dc.rights.accessRightsopen accesses_ES
dc.identifier.doi10.1007/s00604-023-05740-z
dc.relation.projectIDinfo:eu-repo/grantAgreement/MINECO//PID2021-122578NB-I00es_ES
dc.type.hasVersionVoRes_ES


Files in this item

This item appears in the following Collection(s)

Show simple item record

Attribution-NonCommercial-NoDerivatives 4.0 Internacional
This work is under a Creative Commons License Attribution-NonCommercial-NoDerivatives 4.0 Internacional