RT journal article T1 Biomechanical Analysis of Non-Metallic Biomaterial in the Manufacture of a New Knee Prosthesis A1 Suffo Pino, Miguel A1 Revenga, Carlos A2 Ingeniería Mecánica y Diseño Industrial K1 non-metallic knee prosthesis K1 ULTEMTM 1010 biomaterial K1 additive manufacturing K1 biomechanical design K1 two-component knee prosthesis K1 live/dead AB The increase in the number of revision surgeries after a total knee replacement surgeryreaches 19%. One of the reasons for the majority of revisions relates to the debris of the ultra-highmolecular weight polyethylene that serves to facilitate the sliding between the femoral and tibialcomponents. This paper addresses the biomechanical properties of ULTEMTM 1010 in a totally newknee replacement design, based on one of the commercial models of the Stryker manufacturer. It isdesigned and produced through additive manufacturing that replaces the tibial component and thepolyethylene in such a way as to reduce the pieces that are part of the prosthetic assembly to only two:the femoral and the tibial (the so-called “two-component knee prosthesis”). The cytotoxicity as wellas the live/dead tests carried out on a series of biomaterials guarantee the best osteointegration of thestudied material. The finite element simulation method guarantees the stability of the material beforea load of 2000 N is applied in the bending angles 0 , 30 , 60 , 90 , and 120 . Thus, the non-metallicprosthetic material and approach represent a promising alternative for metal-allergic patients. PB MDPI SN 1996-1944 YR 2021 FD 2021-10 LK http://hdl.handle.net/10498/26019 UL http://hdl.handle.net/10498/26019 LA eng NO The live/dead test was exclusively supported by the Andalucia FEDER/ITI 2014-2020 Grant for PI 013/017, as well as the Junta de Andalucia TEP 181 and the CTS 253 PAIDI Spanish research groups. DS Repositorio Institucional de la Universidad de Cádiz RD 16-nov-2026