RT journal article T1 Acoustic conditioning of the neonatal incubator compartment: Improvement proposal A1 Rodríguez Montaño, Víctor Manuel A1 Beira Jiménez, Juan Luis A1 Puyana Romero, Virginia A1 Cueto Ancela, José Luis A1 Hernández Molina, Ricardo A1 Fernández Zacarías, Francisco A2 Ciencias y Técnicas de la Navegación y Construcciones Navales A2 Ingeniería Eléctrica A2 Ingeniería Mecánica y Diseño Industrial A2 Máquinas y Motores Térmicos K1 neonates K1 sound pressure K1 incubators K1 noise source K1 NICUs AB The objective of this work focuses on the study of the main sources of noise associated with incubators and the acoustic effects that derive from them. The method that has been established is based on tests carried out under different operating conditions of the incubators. Noise levels are analyzed under different boundary conditions (Neonatal ICU and "Controlled environment rooms"). Under normal operating conditions, the levels inside the incubator are around 56 dB (A), values that exceed the maximum limits recommended by the American Academy of Pediatrics. The scope of this study is to evaluate the existing noise levels in the incubator and analyze possible design improvements. The study was carried out in the hospitals of Cadiz, Huelva, and Malaga. PB FRONTIERS MEDIA SA SN 2296-2360 YR 2022 FD 2022-09 LK http://hdl.handle.net/10498/28379 UL http://hdl.handle.net/10498/28379 LA eng DS Repositorio Institucional de la Universidad de Cádiz RD 08-nov-2026