Toward In Vivo Transdermal pH Sensing with a Validated Microneedle Membrane Electrode

Identificadores
URI: http://hdl.handle.net/10498/36661
DOI: 10.1021/acssensors.0c02397
ISSN: 2379-3694
Statistics
Metrics and citations
Metadata
Show full item recordDate
2021Department
Química AnalíticaSource
ACS Sensors - 2021, Vol 6, pp. 1129-1137Abstract
We present herein the most complete characterization of
microneedle (MN) potentiometric sensors for pH transdermal measurements
for the time being. Initial in vitro assessment demonstrated suitable analytical
performances (e.g., Nernstian slope, linear range of response from 8.5 to 5.0, and
fast response time) in both buffer media and artificial interstitial fluid (ISF).
Excellent repeatability and reproducibility together with adequate selectivity and
resiliency facilitate the appropriateness of the new pH MN sensor for transdermal
ISF analysis in healthcare. The ability to resist skin insertions was evaluated in
several ex vivo setups using three different animal skins (i.e., chicken, pork, and
rat). The developed pH MN sensor was able to withstand from 5 to 10 repetitive
insertions in all the skins considered with a minimal change in the calibration
graph (<3% variation in both slope and intercept after the insertions). Ex vivo pH
measurements were validated by determining the pH with the MN sensor and a
commercial pH electrode in chicken skin portions previously conditioned at
several pH values, obtaining excellent results with an accuracy of <1% and a precision of <2% in all cases. Finally, pH MN sensors
were applied for the very first time to transdermal measurements in rats together with two innovative validation procedures: (i)
measuring subcutaneous pH directly with a commercial pH microelectrode and (ii) collecting ISF using hollow MNs and then the
pH measurement of the sample with the pH microelectrode. The pH values obtained with pH MN sensors were statistically more
similar to subcutaneous measurements, as inferred by a paired sample t-test at 95% of confidence level. Conveniently, the validation
approaches could be translated to other analytes that are transdermally measured with MN sensors.
Subjects
Ion-selective electrodes; pH sensing; in vivo monitoring; microneedle transdermal sensing; validationCollections