A preparation Penning trap for the TRAPSENSOR project with prospects for MATS at FAIR

Identificadores
URI: http://hdl.handle.net/10498/38791
DOI: 10.1016/j.nimb.2015.11.033
ISSN: 0168-583X
Ficheros
Estadísticas
Métricas y Citas
Metadatos
Mostrar el registro completo del ítemFecha
2016Departamento/s
Física de la Materia CondensadaFuente
Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms, Vol. 376, 2016, pp. 288-291Resumen
Most of the Penning trap spectrometers for precision measurements at radioactive ion beam facilities make use of another Penning trap, located upstream in the experimental set up, to perform isobaric separation and deliver cooled and pure ion samples to be measured. The preparation trap for the project TRAPSENSOR at the University of Granada has been built to prepare ions, produced off-line with a laser-desorption ion source, using firstly the buffer-gas cooling technique. The system has been built following the geometrical specifications given in the Technical Design Report for the MATS Penning-trap system to be built at the future Facility for Antiprotons and Ion Research (FAIR) in Darmstadt. So far, cooling resonances have been obtained for stable nuclides, with mass-to-charge ratios ranging from 40 to about 200, with a performance similar to those systems already in operation at radioactive ion beam facilities. In this contribution, the preparation trap, built and commissioned at the University of Granada, will be briefly described. First results on buffer-gas cooling will be presented.