Electroanalysis of phenylalanine in food supplement by differential pulse voltammetry technique: Application of economical and handmade Sonogel-Carbon electrodes improved with a fast polarization step

Identificadores
URI: http://hdl.handle.net/10498/36859
DOI: 10.1016/j.microc.2025.113208
ISSN: 0026-265X
Statistics
Metrics and citations
Metadata
Show full item recordDate
2025-03-02Department
Química AnalíticaSource
Microchemical Journal, Vol. 212, 2025Abstract
Sonogel-Carbon electrodes have been applied successfully in the determination of phenylalanine (Phe). These handmade economical sensor devices exhibit better electroanalytical performance than commercial electrodes, and by a fast surface polarization step the electrochemical signal was significantly increased. Improvements in the electrochemically active surface area and electrical conductivity were demonstrated. Excellent figures of merits were assessed, such as a sensitivity value of 68.19 ± 3.11 µA mM−1 cm−2 comparable to other previously reported devices and limit of detection of 31.92 ± 3.75 µM. Sonogel-Carbon electrodes exhibit worthy selectivity in the presence of other amino acids and electroactive species. Finally, satisfactory recovery values between 96 and 101 % were obtained when this material was applied in the determination of Phe in a food supplement sample. Sonogel-Carbon electrodes stand out compared to other reported sensor devices due to their simple, low-cost and eco-friendly manufacturing; their reusability; and their robust and selective electrochemical performance.
Subjects
Electrochemical sensor; Sonogel-Carbon; Phenylalanine; Food supplement; VoltammetryCollections
- Artículos Científicos [11954]
- Artículos Científicos IMEYMAT [577]
- Articulos Científicos Quim. Ana. [398]